10 - 5,100 l/min | Model CVF 10-5000-DM
The TrigasDM air flow test stand is a Sonic Nozzle (Critical Nozzle) based air flow measurement system and is the result of more than 30 years of continuously raising the standards of precision and reliability in gas flow measurement.
It has been developed to fully utilise the characteristics of critical nozzles, in particular the highly accurate and repeatable measurement over a wide flow range.
This makes it the ideal solution for accurately measuring airflow through gas turbine engine components for both air and land applications.
Item number: 10571
Maximum Inlet Pressure: | 8 bar g | |
Part Pressure Range: | 0.07 – 1.4 barg (1-20 psig) | |
0.02 – 2 barg (0.25 – 30 psig) optional | ||
Custom ranges available | ||
Part Pressure Ratio Range: | 1.07 – 2.36, 1.02 – 3.04 optional, custom range available | |
Uncertainty: | ±0.3% to ±0.5% of reading or better, depending on calibration accuracy | |
Repeatability: | Better than ±0.15% | |
Cycle Time: | 20 – 60 seconds, capacity dependent | |
Flow Ports Connections: | TriClamp® DN 50 (2”) | |
Pressure Connections: | Swagelok, 6mm (1/4”) | |
Dimensions: | 1,500 x 110 x 1,800 mm (59” x 44” x 71”) Single Manifold (SM) version | |
|
||
Weight: | 350 kg (770 pounds) |
Find here more test stands
CFlow+™ : the software solution from TrigasDM for primary and secondary calibrators
CFlow+™ : the software solution from TrigasDM for primary and secondary calibrators offers a simple and user-friendly way to control and monitor complex test benches. As a Windows® and LabVIEW-based development, CFlow+™ offers intuitive operation of almost all common calibrator systems.
Get in touch with us. We are looking forward to hearing from you.