Lysis™ SMART Pickoffs
New pickoffs which, when combined with our newest Lysis™ Linearizer, vastly simplify flow measurement. A lot of value was placed on making the pickoffs as compact as possible.
SMART-Pickoffs with integrated temperature sensor
Based on the principle of modulated carrier frequency. As there are no magnetic forces involved the rotor is allowed to move freely at lower flow rates. The temperature sensors is built in to the pickoff. It is ideal for viscosity corrections or other applications where temperature also needs to be measured along with the flow rate.
Advantages:
- SMART-Pickoffs contain all the relevant flowmeter data
- Automatic sensor identification (plug & measure)
- The pickoff is „SMART“ enough that it can be used with any Lysis
- TEDS compatible
- Can be installed up to 9 meter between Pickoff and Lysis
- Operating temperature -40 to 125°C
- Specifically designed for applications where limited space is a factor
- Available with two thread variants 5/8″ and 11/16″ for the most common turbines
Product overview Lysis™ LSA-ST-07-V1-07-07-00
Below you will find a product overview of the entire measuring chain including cables with the corresponding item numbers.
with cable 5 meter: item number 101464
with cable 3 meter: item number 12184
with cable 2 meter: item number 12185
with cable 1 meter: item number 10517
lenght 4 meter: item number 10248
lenght 3 meter: item number 12186
lenght 2 meter: item number 12187
A maximum of 1 extension cable can be connected!
LSA-ST-07-V1-07-07-00: item number 11436
ST – SMART-pickoff with temperature sensor, 5 pin
07 – 0-10 V flow + 0-10 V temperature
V1 – 6-36 VDC
07 – Lysis square housing IP68
07 – ODU F G11F1C-P05LJG0/2x ODU G81F1C-P05QJ00
00 – Standard
Lysis V07, 1xBNC (Flow or Temp) + Power: item number 101409
Compatible with any flow meter equipped with a Lysis™ "SMART-Pickoff"
New pickoffs which, when combined with our newest Lysis™ Linearizer, vastly simplify flow measurement. A lot of value was placed on making the pickoffs as compact as possible. SMART Pickoffs with up to 15m cable length files/DM_Bilder/Pickoff_LPP.jpg based on the principle of modulated carrier frequency. As there are no magnetic forces involved the rotor is allowed to move freely at lower flow rates.
After years of experience in flow meter calibration and development of flowmeters, TrigasDM now offers you state-of-the-art flow measurement technology produced in the own factory in Neufahrn. The high-grade Turbine Flowmeters made of stainless steel are characterized by a large linear measuring range and compact design.
Get in touch with us. We are looking forward to hearing from you.